Troubleshooting Guide: Western Blot

Troubleshooting Guide: Western Blot


Problem: Poor Transfer

Possible Source Test or Action
Membrane does not wet uniformly Pre-wet membrane in 100% methanol
MW of protein is less than 10,000 If protein has MW <10,000, it may have 揵lown through?the membrane. Reduce transfer time, or place two sheets of transfer membrane soaked in Anode Buffer II under gel in transfer stack, or select membrane with smaller pore size
Isoelectric point of protein is <9 Use alternate buffer system (such as CAPS buffer pH 10.5) with a higher pH

Assemble transfer stack with additional membrane soaked in cathode buffer on the cathode side of gel. This will capture proteins that migrate towards the cathode. Identify membrane location in stack and immunostain all membranes in stack
Methanol concentration Higher methanol concentration increases binding to membrane but may retard transfer from gel
SDS concentration Adding 0.005 - 0.01% SDS to cathode buffer can increase transfer efficiency of protein from gel
Note: Increased SDS concentration may interfere with binding to membrane
Thick gel Thicker gels or higher MW proteins may require longer transfer times

Problem: High Background Staining

Possible Source Test or Action
Membrane does not wet uniformly Pre-wet membrane in 100% methanol
Inadequate blocking Optimize the blocking step by trying alternate blocking solutions (e.g., non-fat milk, gelatin, etc.) or increasing time and temperature of blocking step
Cross-reactivity of antibody reagents Check for cross-reactivity of antibody reagents to the blocking protein

Problem: Low Protein Binding

Possible Source Test or Action
Overwashing Keep the length of TTBS washes to a minimum

See 揚oor Transfer?problem for alternatives

Problem: High Protein Binding, Low Signal

Possible Source Test or Action
Inadequate antibody staining Check antibody dilutions and expiration dates
Inactive AP conjugate and/or substrate Add enzyme conjugate to substrate reagents as prepared in step 7 of immunostaining procedure. If color occurs, reagents are performing properly
Poor sample If positive control worked, sample may not contain protein of interest or it may be present at concentrations too low to detect

Problem: Several Bands are Stained

Possible Source Test or Action
Non-specific binding of secondary reagents Check non-specific binding by running extra sample lane and omitting 1?antibody from immunostaining procedure, run with a non-specific antibody control in place of 1?antibody, or perform dot blot and omit the 1?antibody. If staining occurs, choose an alternate 2?antibody
Primary antibody is not monospecific Use a monoclonal or affinity purified antibody